Datascope 97E I Datascope 97E I


Availability: listing already sold

Listing description

The new auto-timing supervisor, "R-Trac" is a highly advanced second generation IAB timing control system that combines the sophistication of CardioSync 2 with rapid QRS identification and the fastest IAB inflation and deflation speeds available. The System 97E is the system of choice for handling the most challenging cardiac rhythms.

Faster speeds provide greater diastolic augmentation, enhanced unloading, and a wider use of R-Wave deflation triggering to track and effectively assist patients with arterial fibrillation.

A smart pattern-based timing algorithy, CardioSync™, rapidly adapts to difficult ectopics and sudden rate changes without interruption of pumping.

The 97E also offers a host of user convenience features that make it easier to use in the OR, Cath Lab, CCU and in transport.

Features
Analyses cardiac interval patterns
Automatic selection of R-Wave deflation
Reliable A-Fib tracking
Rapid QRS identification
Improved Arrhythmia Management
SPECIFICATIONS
Mechanical
Dimensions: 43" (H) × 22" (W) × 17" D (109 × 51 × 25 cm)
Off hospital cart: 27" (H) × 50" (W) × 10" (D) (68 × 51 25 cm)
Weight: Console - 82 lbs (37 kg), Battery - 35 lbs (16 kg), Monitor - 8 lbs (4 kg), Hospital Cart - 58 lbs (26 kg)
Electrical
Battery Pack: 24 V DC, 18 Ah
Type: sealed lead acid maintenance free
Battery Operating Time: 2 hours @ 120 bpm
Battery Recharge Time: 18 hours maximum from completely discharge to full
AC Power Requirement: 300 VA normal, 90-130 V, 50/60 Hz; 200-260, 50/60 Hz
Pneumatics
System Compressor: dual head diaphragm pump brushless DC motor
Patient Balloon Pump Gas: medical-grade helium
Blood Detection: automatic protection of compressor pneumatics and isolation of blood in auto fill line
Condensate Removal: automatic condensate removal and disposal via thermo-electric cooling
Display
Type: gas plasma, 5.2" × 8.25" (132 × 210 mm)
Channels: ECG and arterial pressure plus optional balloon pressure waveform
Reference Line: 0-300 mm Hg/max
ECG
Leads: I, II, III, aVR, aVL, aVF, V
Frequency Response: 0.5-135 Hz output jack 0.5 - 12 Hz display
Leakage Current: <10µA source current, <20µA sink current
Defibrillator Protection: discharge levels up to 500 joules
Recovery Time: trace returns to screen, 2 sec maximum
Trigger Source
ECG Trigger Requirement: 120µV±µV at max. gain
Pacer Rejection (no tails): ±2mV to ±700 mV duration 0.1 ms to 2 ms
Pressure: adjustable between 7 mmHg to 30mmHg pulse amplitude
Pacer A-V: A-V interval 80-225 ms rate < or equal to 125 bpm
Pacer A: trigger requirement (same as ECG)
Pacer V: < 185 bpm
Internal: 40-120 in 5 bpm increments; default rate 80 ±1 bpm
ESIS: automatic suppression with internal ECG amplifier
Pressure
Pressure Sensitivity to Trigger: 7mmHg min pulse amplitude
Transducer Requirements: sensitivity - 5.0µV/V/mmHg
Excitation: +5V DC
Zero range: ±120 mmHg
Inputs/Outputs
ECG Connectors: 6-pin 5-lead AAMI proposed standard; patient input 3 - conductor phone jacks; monitor level
Pressure Connectors: 6-pin male connector, Datascope-specified compatible transducer, 3-conductor phone jack; monitor level
Voltage Interface: ECG = 1V/mV nominal arterial pressure = 1V/00 mmHg nominal
Trend Display
Memory: 8 hours total; 30 minutes of 5 minutes average; 7.5 hours of 15 minutes average
Data Available: Heart rate, peak systolic pressure, diastolic pressure, ean and augmented pressure
Recording Format
Recorder: thermal array type recorder/2" thermal paper
Waveform Presentation: ECG, arterial pressure, optional balloon pressure waveform
Graphic Presentation: heart rate trend, pressure trends
Vent Marker: trigger point.
Inflate interval
Annotations
ECG: lead
Pressure: scale
Parameters: heart rate, systolic, diastolic, mean augmented pressure
Trend: same as trend display

This equipment is located in US


Please note that this description may have been translated automatically.

Availability: listing already sold

Listing information

Manufacturer Datascope 97E I
Model Datascope 97E I
Year 2015
Country USA USA
Condition Good
Main category Lab, Medical and Bioscience
Subcategory Hospital Equipment
ID P81114079

Availability: listing already sold

About the seller

Client type Machinery dealer
On Kitmondo since 2018
Number of listings 0
Country USA USA
Last activity Nov. 21, 2018